Module Documentation¶
fluke5440b_async¶
This is an asyncIO library for the Fluke 5440B calibrator. It manages all functions of the calibrator and takes care of the internal state.
- class fluke5440b_async.Fluke_5440B(connection: AsyncGpib | AsyncPrologixGpibController, log_level: int = 30)[source]¶
The driver for the Fluke 5440B voltage calibrator. It supports both linux-gpib and the Prologix GPIB adapters.
- __init__(connection: AsyncGpib | AsyncPrologixGpibController, log_level: int = 30)[source]¶
Create Fluke 5440B device with the GPIB connection given.
- Parameters:
connection (AsyncGpib or AsyncPrologixGpibController) – The GPIB connection.
log_level (int, default=logging.WARNING) – The level of logging output. By default, only warnings or higher are output.
- async acal() None [source]¶
Run the internal calibration routine. It will take about 6.5 minutes during which the instrument hardware is blocked. See page 3-2 of the operator manual for details.
- async connect() None [source]¶
Connect the GPIB connection and configure the GPIB device for the DMM. This function must be called from the AsyncIO loop and takes care of connecting the GPIB adapter.
- property connection: AsyncGpib | AsyncPrologixGpibController¶
The GPIB connection.
- async disconnect() None [source]¶
Disconnect from the instrument and clean up. This call will also automatically remove the local lockout if set.
- async get_calibration_constants() CalibrationConstants [source]¶
Query the calibration constants and gain shifts with respect to the previous internal calibration. See page 3-18 of the operator manual for details. :returns: A dataclass containing the constants as Decimals :rtype: CalibrationConstants
- async get_current_limit() Decimal | tuple[Decimal, Decimal] [source]¶
Get the current limit set on the instrument. It will raise an error, when in voltage boost mode. :returns: The positive and negative limit :rtype: tuple of Decimal
- Raises:
DeviceError – Raised if there was an error processing the command.
- async get_error() int [source]¶
Get the last error thrown by the instrument if any. It is recommended to check for errors after using the
write()
function, if the test_error parameter is not set. :returns: The last error thrown. It might be an ErrorCode or a SelfTestErrorCode. This is ambiguous and depends on thelast command.
- Return type:
int
- async get_id() tuple[str, str, str, str] [source]¶
Returns the instrument name and the software version string. To emulate the *IDN? SCPI command, the result is a tuple containing the Manufacturer, Device, Serial (0) and FW version. :returns: A tuple containing the manufacturer id, device id, a zero, and the software version. :rtype: tuple of str
- async get_output() Decimal [source]¶
Returns the output voltage currently set. :returns: The output voltage set. :rtype: Decimal
- Raises:
DeviceError – Raised if there was an error processing the command.
- async get_rs232_baud_rate() int | float [source]¶
Return the RS-232 the baud rate in bit/s. :returns: The baud rate in bit/s :rtype: float
- Raises:
DeviceError – If test_error is set to True and there was an error processing the command.
- async get_separator() SeparatorType [source]¶
Returns the separator used by the instrument to separate multiple queries. :returns: The separator used by the device. :rtype: SeparatorType
- Raises:
DeviceError – Raised if there was an error processing the command.
- async get_srq_mask() SrqMask [source]¶
Get the SRQ mask. This mask is used to determine, when the device will signal the SRQ line. :returns: The bitmask to determine the bits to signal. :rtype: SrqMask
- Raises:
DeviceError – If test_error is set to True and there was an error processing the command.
- async get_state() DeviceState [source]¶
While the instrument is running long jobs, it signals its current state. Use this to poll the state. :returns: The current device state. :rtype: DeviceState
- Raises:
DeviceError – Raised if there was an error processing the command.
- async get_status() StatusFlags [source]¶
Get the status of the instrument. The status flags contain the mode the instrument is running in, like boost mode, or the state of the external sense connection, etc. :returns: The status flags of the different settings :rtype: StatusFlags
- Raises:
DeviceError – Raised if there was an error processing the command.
- async get_terminator() TerminatorType [source]¶
Returns the line terminator sent by the instrument. :returns: The line terminator used by the device. :rtype: TerminatorType
- async get_voltage_limit() tuple[Decimal, Decimal] [source]¶
Get the voltage limit set on the instrument. It will raise an error, when in current boost mode. :returns: The positive and negative limit :rtype: tuple of Decimal
- Raises:
DeviceError – Raised if there was an error processing the command.
- async local() None [source]¶
Enable the front panel buttons, if the instrument is in local lock out.
- async query(cmd: str | bytes, test_error: bool = False) str | list[str] [source]¶
Write a string or bytestring to the instrument, then immediately read back the result. This is a combined call to
write()
andread()
. :param cmd: The command written to the device :type cmd: str or bytes :param test_error: Check for errors by serial polling after sending the command. :type test_error: bool, default=False- Returns:
Returns either a simple string, or if multiple results are returned, a list of strings.
- Return type:
str or list of str
- Raises:
DeviceError – If test_error is set to True and there was an error processing the command.
- async read() str | list[str] [source]¶
Read from the device. :returns: Returns either a simple string, or if multiple results are returned, a list of strings. :rtype: str or list of str
- async reset() None [source]¶
Place the instrument in standby, enable voltage mode, set the output voltage to 0.0, disable the divider output, the external guard mode and external sense mode. :raises DeviceError: Raised if there was an error processing the command.
- async selftest_all() None [source]¶
Run all three self-tests. This function is a combination of
selftest_digital()
,selftest_analog()
andselftest_hv()
.- Raises:
DeviceError – Raised if there was an error processing the command.
- async selftest_analog() None [source]¶
Test the ADC, the low voltage part and the oven. It will take about 4 minutes during which the instrument hardware is blocked. See page 3-19 of the operator manual for details.
- Raises:
DeviceError – Raised if there was an error processing the command.
- async selftest_digital() None [source]¶
Test the main CPU, the front panel CPU and the guard CPU. It will take about 5 seconds during which the instrument hardware is blocked. See page 3-19 of the operator manual for details.
- Raises:
DeviceError – Raised if there was an error processing the command.
- async selftest_hv() None [source]¶
Test the ADC and the high voltage part. It will take about 1 minute during which the instrument hardware is blocked. See page 3-20 of the operator manual for details.
- Raises:
DeviceError – Raised if there was an error processing the command.
- async serial_poll() SerialPollFlags [source]¶
Poll the serial output buffer of the device. This can be used to query for the SRQ bit when device requests service, has encountered an error or is busy. :returns: The content of the serial output buffer :rtype: SerialPollFlags
- async set_current_limit(value: int | float | Decimal, value2: int | float | Decimal | None = None) None [source]¶
Set the positive and negative current limit. :param value: Either a positive or negative number to set the positive or negative limit. :type value: int or float or Decimal :param value2: Either a positive or negative number to set the positive or negative limit. If omitted, only one limit will
be set.
- Raises:
DeviceError – Raised if there was an error processing the command.
ValueError – Raised if the limits are out of bounds.
- async set_divider(enabled: bool) None [source]¶
Enable the internal 1:10 and 1:100 divider to reduce the output noise and increase the resolution of voltages in the range -2.2 V to 2.2 V. Do not enable the external sense connection via
set_internal_sense(False)()
, as this will decrease the accuracy. The divider has an output impedance of 450 Ω. The load should ideally be greater than 1 GΩ to keep the loading error below 1 ppm. See page 3-10 of the operator manual for details.- Parameters:
enabled (bool) – Set to True to enable the divider.
- Raises:
DeviceError – Raised if there was an error processing the command.
- async set_enable_rs232(enabled: bool) None [source]¶
Enable the RS-232 printer port. :param enabled: True to enable the RS-232 port. :type enabled: bool
- Raises:
DeviceError – If test_error is set to True and there was an error processing the command.
- async set_internal_guard(enabled: bool) None [source]¶
If set, the guard is internally connected to the output LO terminal. Use this, if the device being calibrated has floating inputs. If calibrating devices with grounded inputs, connect the guard terminal to the input LO of the device and disable the internal guard. See page 2-14 of the operator manual for details. :param enabled: Set to True for floating devices. :type enabled: bool
- Raises:
DeviceError – Raised if there was an error processing the command.
- async set_internal_sense(enabled: bool) None [source]¶
If the load resistance is greater than 1 MΩ, 2-wire calibration can be used. Otherwise, the cable resistance will reduce the accuracy. Use internal sense for 2-wire calibrations. See page 2-13 of the operator manual for details.
- Parameters:
enabled (bool) – Set to True for 2-wire sense.
- Raises:
DeviceError – Raised if there was an error processing the command.
- set_log_level(loglevel: int = 30)[source]¶
Set the log level of the library. By default, the level is set to warning. :param loglevel: The log level of the library :type loglevel: int, default=logging.WARNING
- async set_mode(value: ModeType) None [source]¶
Enabled either voltage or current boost mode using an external Fluke 5205A power amplifier or a Fluke 5220A transconductance amplifier. :param value: Either normal mode, current or voltage boost. :type value: ModeType
- Raises:
DeviceError – Raised if there was an error processing the command.
- async set_output(value: int | float | Decimal, test_error: bool = True) None [source]¶
Set the output of the calibrator. If an output greater than ±22 V is set, the calibrator will automatically go to STBY for safety reasons. Call set_output_enabled(True) to re-enable the output. :param value: The value to be set :type value: int or float or Decimal :param test_error: Raise an exception if there was an error :type test_error: bool, Default=True
- Raises:
DeviceError – Raised if there was an error processing the command and test_error was set to True.
- async set_output_enabled(enabled: bool) None [source]¶
Set the output to either STANDBY or OPR. :param enabled: Set to OPR if true :type enabled: bool
- Raises:
DeviceError – Raised if there was an error processing the command.
- async set_rs232_baud_rate(value: int | float) None [source]¶
Set the baud rate of the RS-232 interface. :param value: The baud rate. :type value: {50, 75, 110, 134.5, 150, 200, 300, 600, 1200, 1800, 2400, 4800, 9600}
- Raises:
DeviceError – If test_error is set to True and there was an error processing the command.
- async set_srq_mask(value: SrqMask) None [source]¶
Set the service request mask register. Each bit set, will signal the SRQ line, when a service request of the device is triggered. :param value: The bitmask to set. :type value: SrqMask
- Raises:
DeviceError – If test_error is set to True and there was an error processing the command.
- async set_voltage_limit(value: int | float | Decimal, value2: int | float | Decimal | None = None) None [source]¶
Set the positive and negative voltage limit. :param value: Either a positive or negative number to set the positive or negative limit. :type value: int or float or Decimal :param value2: Either a positive or negative number to set the positive or negative limit. If omitted, only one limit will
be set.
- Raises:
DeviceError – Raised if there was an error processing the command.
ValueError – Raised if the limits are out of bounds.
- async write(cmd: str | bytes, test_error: bool = False)[source]¶
Write a string or bytestring to the instrument. :param cmd: The command written to the device :type cmd: str or bytes :param test_error: Check for errors by serial polling after sending the command. :type test_error: bool, default=False
- Raises:
DeviceError – If test_error is set to True and there was an error processing the command.
Dataclasses¶
- class fluke5440b_async.CalibrationConstants(gain_02V: Decimal, gain_2V: Decimal, gain_10V: Decimal, gain_20V: Decimal, gain_250V: Decimal, gain_1000V: Decimal, offset_10V_pos: Decimal, offset_20V_pos: Decimal, offset_250V_pos: Decimal, offset_1000V_pos: Decimal, offset_10V_neg: Decimal, offset_20V_neg: Decimal, offset_250V_neg: Decimal, offset_1000V_neg: Decimal, gain_shift_10V: Decimal, gain_shift_20V: Decimal, gain_shift_250V: Decimal, gain_shift_1000V: Decimal, resolution_ratio: Decimal, adc_gain: Decimal)[source]¶
The calibration constants of the Fluke 5440B.
- adc_gain: Decimal¶
- gain_02V: Decimal¶
- gain_1000V: Decimal¶
- gain_10V: Decimal¶
- gain_20V: Decimal¶
- gain_250V: Decimal¶
- gain_2V: Decimal¶
- gain_shift_1000V: Decimal¶
- gain_shift_10V: Decimal¶
- gain_shift_20V: Decimal¶
- gain_shift_250V: Decimal¶
- offset_1000V_neg: Decimal¶
- offset_1000V_pos: Decimal¶
- offset_10V_neg: Decimal¶
- offset_10V_pos: Decimal¶
- offset_20V_neg: Decimal¶
- offset_20V_pos: Decimal¶
- offset_250V_neg: Decimal¶
- offset_250V_pos: Decimal¶
- resolution_ratio: Decimal¶
Enums and Flags¶
Flags are used for the status registers returned by the device.
- class fluke5440b_async.flags.SerialPollFlags(value, names=None, *values, module=None, qualname=None, type=None, start=1, boundary=None)[source]¶
The register returned by a serial poll.
- DOING_STATE_CHANGE = 4¶
- ERROR_CONDITION = 32¶
- MSG_RDY = 8¶
- NONE = 0¶
- OUTPUT_SETTLED = 16¶
- SRQ = 64¶
- class fluke5440b_async.flags.SrqMask(value, names=None, *values, module=None, qualname=None, type=None, start=1, boundary=None)[source]¶
The register to control the service request interrupts.
- DOING_STATE_CHANGE = 4¶
- ERROR_CONDITION = 32¶
- MSG_RDY = 8¶
- NONE = 0¶
- OUTPUT_SETTLED = 16¶
- class fluke5440b_async.flags.StatusFlags(value, names=None, *values, module=None, qualname=None, type=None, start=1, boundary=None)[source]¶
The internal status register that holds the device configuration.
- CURRENT_BOOST_MODE = 2¶
- DIVIDER_ENABLED = 8¶
- INTERNAL_GUARD_ENABLED = 64¶
- INTERNAL_SENSE_ENABLED = 16¶
- OUTPUT_ENABLED = 32¶
- REAR_OUTPUT_ENABLED = 128¶
- VOLTAGE_BOOST_MODE = 4¶
- VOLTAGE_MODE = 1¶
Enums are used to represent the device functions and settings.
- class fluke5440b_async.enums.DeviceState(value, names=None, *values, module=None, qualname=None, type=None, start=1, boundary=None)[source]¶
The internal device state.
- CALIBRATING_ADC = 16¶
- CALIBRATING_GAIN_10V_NEG = 53¶
- CALIBRATING_GAIN_10V_POS = 48¶
- CALIBRATING_GAIN_20V_NEG = 52¶
- CALIBRATING_GAIN_20V_POS = 49¶
- CALIBRATING_GAIN_HV_NEG = 51¶
- CALIBRATING_GAIN_HV_POS = 50¶
- CAL_N1_N2_RATIO = 96¶
- EXT_CAL_02V = 69¶
- EXT_CAL_02V_NULL = 85¶
- EXT_CAL_1000V = 67¶
- EXT_CAL_1000V_NULL = 83¶
- EXT_CAL_10V = 64¶
- EXT_CAL_10V_NULL = 80¶
- EXT_CAL_20V = 65¶
- EXT_CAL_20V_NULL = 81¶
- EXT_CAL_250V = 66¶
- EXT_CAL_250V_NULL = 82¶
- EXT_CAL_2V = 68¶
- EXT_CAL_2V_NULL = 84¶
- IDLE = 0¶
- PRINTING = 208¶
- RESETTING = 240¶
- SELF_TEST_FRONTPANEL_CPU = 113¶
- SELF_TEST_GUARD_CPU = 114¶
- SELF_TEST_HIGH_VOLTAGE = 129¶
- SELF_TEST_LOW_VOLTAGE = 128¶
- SELF_TEST_MAIN_CPU = 112¶
- SELF_TEST_OVEN = 130¶
- WRITING_TO_NVRAM = 224¶
- ZEROING_1000V_NEG = 39¶
- ZEROING_1000V_POS = 38¶
- ZEROING_10V_NEG = 33¶
- ZEROING_10V_POS = 32¶
- ZEROING_20V_NEG = 35¶
- ZEROING_20V_POS = 34¶
- ZEROING_250V_NEG = 37¶
- ZEROING_250V_POS = 36¶
- class fluke5440b_async.enums.ErrorCode(value, names=None, *values, module=None, qualname=None, type=None, start=1, boundary=None)[source]¶
The error codes used by the Fluke 5440B.
- BOOST_INTERFACE_CONNECTION_ERROR = 144¶
- BOOST_INTERFACE_CURRENT_TRIP = 147¶
- BOOST_INTERFACE_MISSING = 145¶
- BOOST_INTERFACE_VOLTAGE_TRIP = 146¶
- BUFFER_OVERFLOW = 157¶
- DIVIDER_OUT_OF_RANGE = 171¶
- GPIB_HANDSHAKE_ERROR = 152¶
- INVALID_CHARACTER = 158¶
- INVALID_COMMAND = 175¶
- INVALID_GUARD_MODE = 173¶
- INVALID_PARAMETER = 156¶
- INVALID_SENSE_MODE = 172¶
- LIMIT_OUT_OF_RANGE = 170¶
- NONE = 0¶
- OUTPUT_OUTSIDE_LIMITS = 169¶
- PARAMETER_OUT_OF_RANGE = 168¶
- RS232_ERROR = 160¶
- SEPARATOR_ERROR = 154¶
- TERMINATOR_ERROR = 153¶
- UNKNOWN_COMMAND = 155¶
- class fluke5440b_async.enums.ModeType(value, names=None, *values, module=None, qualname=None, type=None, start=1, boundary=None)[source]¶
The output modes. Voltage boost means a connected Fluke 5205A power amplifier and current boost means a connected Fluke 5220A transconductance amplifier.
- CURRENT_BOOST = 'BSTC'¶
- NORMAL = 'BSTO'¶
- VOLTAGE_BOOST = 'BSTV'¶
- class fluke5440b_async.enums.SelfTestErrorCode(value, names=None, *values, module=None, qualname=None, type=None, start=1, boundary=None)[source]¶
The error codes thrown by the self-test routine. See page 4-8 of the service manual for details.
- ANALOG_MEASURE_FAULT_CHECK_ANALOG_BUSS = 72¶
- ANALOG_MEASURE_FAULT_CHECK_ZERO_AMP = 73¶
- ANALOG_MEASURE_FAULT_GAIN_SHIFT_TOO_LARGE = 75¶
- ANALOG_MEASURE_FAULT_UNABLE_TO_ZERO_RANGE = 74¶
- BOARD_ACK_FAULT_CHECK_DAC_BOARD = 48¶
- BOARD_ACK_FAULT_CHECK_OUTPUT_BOARD = 51¶
- BOARD_ACK_FAULT_CHECK_PREAMP_BOARD = 49¶
- BOARD_ACK_FAULT_CHECK_SAMPLE_BOARD = 50¶
- BOOST_INTERFACE_ERROR_CHECK_MISSING_REAR_CABLE = 145¶
- BOOST_INTERFACE_ERROR_CHECK_REAR_CONNECTOR = 144¶
- BOOST_INTERFACE_ERROR_CURRENT_TRIP = 147¶
- BOOST_INTERFACE_ERROR_VOLTAGE_TRIP = 146¶
- CAL_SWITCH_LOCKED = 174¶
- DAC_ANALOG_FAULT_CHECK_0V_OUTPUT = 87¶
- DAC_ANALOG_FAULT_CHECK_10V_OUTPUT = 90¶
- DAC_ANALOG_FAULT_CHECK_5V_DAC_CKT = 93¶
- DAC_ANALOG_FAULT_CHECK_5V_DAC_REC = 96¶
- DAC_ANALOG_FAULT_CHECK_DAC_FILTER = 92¶
- DAC_ANALOG_FAULT_CHECK_DAC_OVEN = 95¶
- DAC_ANALOG_FAULT_CHECK_NEG_10V_OUTPUT = 91¶
- DAC_ANALOG_FAULT_CHECK_NEG_5V_REGULATOR = 89¶
- DAC_ANALOG_FAULT_CHECK_REFERENCE = 88¶
- DAC_ANALOG_FAULT_CHECK_SECOND_SPEED = 94¶
- DAC_DIGITAL_FAULT_BIAS_SIGNAL = 83¶
- DAC_DIGITAL_FAULT_CHECK_A_TO_D = 80¶
- DAC_DIGITAL_FAULT_CHECK_FIRST_SWITCH = 81¶
- DAC_DIGITAL_FAULT_SECOND_SWITCH = 82¶
- FIL_BOUT_BRDS_FAULT_CHECK_1100V_RANGE = 132¶
- FIL_BOUT_BRDS_FAULT_CHECK_275V_RANGE = 129¶
- FIL_BOUT_BRDS_FAULT_CHECK_550V_RANGE = 130¶
- FIL_BOUT_BRDS_FAULT_CHECK_875V_RANGE = 131¶
- FIL_BOUT_BRDS_FAULT_CHECK_NEG_275V_RANGE = 128¶
- FRONT_COMMUNICATION_FAULT = 64¶
- FRONT_DIGITAL_FAULT_MEMORY = 32¶
- FRONT_DIGITAL_FAULT_PROCESSOR = 33¶
- GRD_PWR_SUPPLY_FAULT_CHECK_20V_OVEN = 97¶
- GUARD_COMMUNICATION_FAULT_CHECK_GARBLED_DATA = 56¶
- GUARD_COMMUNICATION_FAULT_GUARD_NOT_ANSWERING = 57¶
- IEEE488_REMOTE_ERROR_EXPECTING_BAD_CHARACTER = 158¶
- IEEE488_REMOTE_ERROR_EXPECTING_BUFFER_OVERFLOW = 157¶
- IEEE488_REMOTE_ERROR_EXPECTING_HEADER = 155¶
- IEEE488_REMOTE_ERROR_EXPECTING_NUMBER = 156¶
- IEEE488_REMOTE_ERROR_EXPECTING_SEPARATOR = 154¶
- IEEE488_REMOTE_ERROR_EXPECTING_TERMINATOR = 153¶
- IEEE488_REMOTE_ERROR_SOURCE_HANDSHAKE = 152¶
- INSIDE_GUARD_FAULT_CHECK_ADDRESS_BUS = 42¶
- INSIDE_GUARD_FAULT_CHECK_DATA_BUS = 41¶
- INSIDE_GUARD_FAULT_CHECK_GUARD_CONTROL_BUS = 43¶
- INSIDE_GUARD_FAULT_CHECK_GUARD_MEMORY = 40¶
- MAIN_CONTROL_FAULT_CHECK_MAIN_CONTROLLER = 16¶
- MAIN_CONTROL_FAULT_CHECK_MAIN_MEMORY = 17¶
- MAIN_CONTROL_FAULT_CHECK_MAIN_NV_MEMORY = 18¶
- MAIN_INTERRUPT_FAULT_CHECK_INPUT_FR_FRONT = 27¶
- MAIN_INTERRUPT_FAULT_CHECK_INPUT_FR_GUARD = 30¶
- MAIN_INTERRUPT_FAULT_CHECK_MAIN_CLOCK = 28¶
- MAIN_INTERRUPT_FAULT_CHECK_NVMEMORY_TIMER = 25¶
- MAIN_INTERRUPT_FAULT_CHECK_REMOTE_INPUT = 29¶
- MAIN_INTERRUPT_FAULT_CHECK_SERIAL_INPUT = 26¶
- MAIN_INTERRUPT_FAULT_CHECK_SERIAL_OUTPUT_TIMER = 24¶
- NONE = 0¶
- OUTPUT_BOARDS_FAULT_CHECK_02V_RANGE = 115¶
- OUTPUT_BOARDS_FAULT_CHECK_2V_RANGE = 113¶
- OUTPUT_BOARDS_FAULT_CHECK_CURR_LIM_CKT = 114¶
- OUTPUT_BOARDS_FAULT_CHECK_ZERO_AMP = 112¶
- OUTPUT_LIMIT_FAULT = 136¶
- OUTPUT_LIMIT_FAULT_OUTPUT_OVER_VOLTAGE = 137¶
- OUTPUT_LIMIT_FAULT_OUTPUT_UNDER_VOLTAGE = 138¶
- POWER_SUPPLY_FAULT_CHECK_GUARDED_POWER = 9¶
- POWER_SUPPLY_FAULT_TEST_UNGUARDED_POWER = 8¶
- PREAMP_ANALOG_FAULT_CHECK_INTCAL_CONFIG = 104¶
- PREAMP_ANALOG_FAULT_CHECK_PREAMP_OVEN = 110¶
- PREAMP_OUT_BRDS_FAULT_CHECK_250V_RANGE = 116¶
- PREAMP_OUT_BRDS_FAULT_CHECK_STANDBY_CONFIG = 109¶
- PREAMP_SAMPLE_STRING_FAULT_CHECK_HIGH_V_INTCAL = 123¶
- RS232C_SERIAL_ERROR = 160¶
- SAMPLE_STRING_FAULT_CHECK_10V_INTCAL = 120¶
- SAMPLE_STRING_FAULT_CHECK_1KV_RANGE = 122¶
- SAMPLE_STRING_FAULT_CHECK_20V_INTCAL = 121¶
- USER_ENTRY_ERROR_DIVIDER_OUT_OF_RANGE = 171¶
- USER_ENTRY_ERROR_INSTRUMENT_IS_BUSY = 175¶
- USER_ENTRY_ERROR_IN_OUTPUT_TERMINAL_GUARD = 173¶
- USER_ENTRY_ERROR_IN_OUTPUT_TERMINAL_SENSE = 172¶
- USER_ENTRY_ERROR_LIMITS_OUT_OF_RANGE = 170¶
- USER_ENTRY_ERROR_NUMBER_OUT_OF_RANGE = 168¶
- USER_ENTRY_ERROR_OUTPUT_OUT_OF_RANGE = 169¶
Errors¶
Custom errors raised by the Fluke 5440B.
- exception fluke5440b_async.errors.DeviceError(message: str, error_code: ErrorCode)[source]¶
The device returned an error during operation.
- exception fluke5440b_async.errors.Fluke5440bError[source]¶
The base class for all Fluke 5540B exceptions
- exception fluke5440b_async.errors.SelftestError(selftest: str, error_code: SelfTestErrorCode)[source]¶
The device returned an error during self-test. See page 4-8 of the service manual for details.